Acetone CH3COCH3 NO.151L


Performance

Measuring Range

50 to 4000 ppm

4000 to 12000 ppm

Number of Pump Strokes

2

1

Correction Factor

1

3

Sampling Time

2 minutes per pump stroke

Detecting Limit

5 ppm (n=2)

Color Change

Yellow Red

Reaction Principle

CH3COCH3 + (NH2OH)3EH3PO4 H3PO4+ Base
Reddish product

Coefficient of Variation 15% (for 50 to 500 ppm), 10% (for 500 to 4000 ppm)
Shelf Life 2 Years
Corrections for
temperature & humidity
Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10oC (50oF) or below.

Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Acrolein, Acetaldehyde

1/10 time

Plus error

Produce reddish stain

Aromatic hydrocarbons

-

No effect

No discoloration

Methyl isobutyl ketone

-

Plus error

Produce reddish stain

Methyl ethyl ketone

-

Plus error

Produce reddish stain

Other substance measurable with this detector tube

Substance

Correction Factor

Pump Strokes

Measuring Range

Methyl ethyl ketone

Factor: 0.42

5

21 to 1680 ppm

Propionaldehyde

Factor: 0.47

2

24 to 1880 ppm



Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
500ppm 750ppm 2.1 to 13%



Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.