Aniline C6H5NH2 NO.181


Performance

Measuring Range

1.25 to 2.5 ppm

2.5 to 30 ppm

30 to 60 ppm

Number of Pump Strokes

5

3

2

Correction Factor

1/2

1

2

Sampling Time

3 minutes per pump stroke

Detecting Limit

0.25 ppm (n=5)

Color Change

Pale yellow Pale green

Reaction Principle

Aniline reduces sodium dichromate to form phosphate, which is pale green in color.

Coefficient of Variation 10% (for 2.5 to 10 ppm), 5% (for 10 to 30 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Ammonia

1/10 time

Plus error

No discoloration

Amines

1/10 time

Plus error

No discoloration

Aromatic amines

-

Plus error

Produce pale green discoloraiton


Other substance measurable with this detector tube

Substance

Correction

No. of pump strokes

Measuring range

N-methylaniline

Factor:1.4

2

3.5 to 42 ppm

N,N-Dimethylaniline

Factor:1.0

3

2.5 to 30 ppm

o-Toluidine

Factor: 2.0

2

5 to 60 ppm


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
2ppm - 1.3 to 11%


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.