| Acid gases | NO.80 | ||
|
|
|||

| Performance | (Acetic acid) |
|
Measuring Range |
1 to 2 ppm |
2 to 40 ppm |
40 to 80 ppm |
|
Number of Pump Strokes |
4 |
2 |
1 |
|
Correction Factor |
1/2 |
1 |
2 |
|
Sampling Time |
45 seconds per pump stroke |
||
|
Detecting Limit |
0.5 ppm (n=4) |
||
|
Color Change |
Pale reddish purple |
||
|
Reaction Principle |
Acid gases react with detecting layer to produce yellow stain. |
||
| Coefficient of Variation | 10% (for 2 to 40 ppm) | ||
| Shelf Life | 2 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes in dark and cool place | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Correction |
Pump |
Color Change |
Measuring range |
|
Hydrogen chloride |
4 |
2 |
Pale reddish purple |
4 to 320 ppm |
|
Chlorine |
0.35 |
2 |
White |
0.35 to 28 ppm |
|
Nitric acid |
2.5 |
2 |
Pale reddish purple |
2.5 to 200 ppm |
|
Sulfur dioxide |
0.75 |
2 |
Yellow |
0.75 to 60 ppm |
|
Nitrogen dioxide |
0.1 |
2 |
Pinkish gray |
0.1 to 80 ppm |
|
Iodine |
0.06 |
2 |
Light bluish gray |
0.06 to 48 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| - | - | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |