| Acetic acid | CH3CO2H | NO.81 | |
|
|
|||

| Performance |
|
Measuring Range |
1 to 2 ppm |
2 to 50 ppm |
50 to 100 ppm |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2 |
|
Sampling Time |
1 minute per pump stroke |
||
|
Detecting Limit |
0.2 ppm (n=2) |
||
|
Color Change |
Pink |
||
|
Reaction Principle |
Acetic acid neutralizes sodium hydroxide to discolor indicator to yellow. |
||
| Coefficient of Variation | 10% (for 2 to 10ppm), 5% (for 10 to 50 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Humidity correction is necessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Hydrogen chloride, Hydrogen cyanide,Nitric acid |
3 times or higher |
Plus error |
Produce yellow stain |
|
Chlorine, Sulfur dioxide, Nitrogen dioxide |
1/2 times or higher |
Plus error |
Produce yellow stain |
| Other substance measurable with this detector tube |
|
Substance |
Correction |
No. of pump strokes |
Measuring range |
|
Acetic anhydride |
Factor: 0.3 |
1 |
0.6 to 15 ppm |
|
Acrylic acid |
Factor: 1.0 |
1 |
2 to 50 ppm |
|
Formic acid |
Factor: 2.6 |
1 |
5.2 to 130 ppm |
| Isovalenic acid | Factor: 1.0 | 1 | 2 to 50 ppm |
|
Maleic anhydride |
Factor: 0.4 |
1 |
0.8to 20 ppm |
|
Methacrylic acid |
Factor: 0.9 |
1 |
1.8to 45 ppm |
|
Propionic acid |
Factor: 1.5 |
1 |
3to 75 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 10ppm | 15ppm | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |