Benzene C6H6 NO.121L

121L


Performance

Measuring Range

0.125 to 0.25 ppm

0.25 to 12 ppm

12 to 60 ppm

Number of Pump Strokes

10

5

1

Correction Factor

1/2

1

5

Sampling Time

3 minutes per pump stroke

Detecting Limit

0.1ppm (n=10)

Color Change

White Brownish gray

Reaction Principle

Benzene reduces iodine pentoxide to liberate iodine,
which produces a brownish gray in color

Coefficient of Variation 10% (for 0.25 to 4 ppm), 5% (for 4 to 12 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Aromatic HCs

-

Plus error

Produce brownish gray stain


Other substance measurable with this detector tube

Substance

Correction Factor

Pump Strokes

Measuring Range

Methylene Iodide

1.67

5

0.4 to 20 ppm

Methyl Iodide

2.8

5

0.7 to 33.6 ppm


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
0.5ppm 2.5ppm 1.3 to 7.1%


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.