| Butyl acetate | CH3CO2(CH2)3CH3 (CH3CO2C4H9) |
NO.142L | |
|
|
|||

| Performance |
|
Measuring Range |
10 to 300 ppm |
|
Number of Pump Strokes |
2 |
|
Correction Factor |
1 |
|
Sampling Time |
4 minutes per pump stroke |
|
Detecting Limit |
2 ppm (n=2) |
|
Color Change |
Yellow |
|
Reaction Principle |
CH3CO2(CH2)3CH3 + Cr6+ + H2SO4 |
| Coefficient of Variation | 15% (for 10 to 100 ppm), 10% (for 100 to 300 ppm) |
| Shelf Life | 2 Years |
| Corrections for temperature & humidity |
Temperature correction is necessary |
| Store the tubes at cool and dark place. | |
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Alcohols |
- |
Plus error |
Dark brown (few minutes later) |
|
Ketones |
- |
Plus error |
|
|
Esters |
- |
Plus error |
|
|
Aromatic HCs |
- |
Plus error |
| Other substance measurable with this detector tube |
|
Substance |
Correction |
No. of Pump Strokes |
Measuring Range |
|
Isobuty Acrylate |
Factor: 0.26 |
2 |
2.6 to 78 ppm |
|
Butyl Acrylate |
Factor 0.7 |
2 |
7 to 210 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 150ppm | 200ppm | 1.7 to 7.6% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |