| 1,3-Butadiene | CH2:CHCH:CH2 | NO.174L |
|
|
||

| Performance |
|
Measuring Range |
2.5 to 5 ppm |
5 to 100 ppm |
|
Number of Pump Strokes |
8 |
4 |
|
Correction Factor |
1/2 |
1 |
|
Sampling Time |
3 minutes per pump stroke |
|
|
Detecting Limit |
0.5 ppm (n=8) |
|
|
Color Change |
Pale yellow |
|
|
Reaction Principle |
1,3-Butadiene reacts with palladium sulfate and |
|
| Coefficient of Variation | 10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm) | |
| Shelf Life | 3 Years | |
| Corrections for temperature & humidity |
Temperature correction is necessary | |
| Store the tubes at cool and dark place. | ||
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Acetylene, Hydrogen cyanide |
- |
Plus error |
Discolor to white |
|
Carbon monoxide, Hydrogen |
- |
Plus error |
Discolors to blue for whole layer |
|
Hydrogen chloride |
- |
Plus error |
Discolors to pink stain |
|
Hydrogen sulfide |
- |
Plus error |
Discolors to black |
|
Ethylene, Propylene, Butylene |
- |
Plus error |
Discolors to blue |
|
Butane, Pentane |
- |
Plus error |
Discolor to blue for whole layer |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
1,3-Pentadiene |
8.5 |
4 |
42.5 to 850 ppm |
| Calibration gas generation | Permeation tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 2ppm | - | 2.0 to 12% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |