1,3-Butadiene CH2:CHCH:CH2 NO.174LL

174LL


Performance

Measuring Range

0.5 to 5 ppm

Number of Pump Strokes

1

Correction Factor

1

Sampling Time

3 minutes per pump stroke

Detecting Limit

0.1 ppm (n=1)

Color Change

Pink Pale Yellow

Reaction Principle

CH2:CHCH:CH2+ KMnO4 Reaction Product

Coefficient of Variation 10% (for 0.05 to 1 ppm-hr), 5% (for 2 to 5 ppm-hr)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes in the refrigerator to keep at 10oC (50oF) or below.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Ammonia

-

No effect

No discoloration

CO & CO2

-

No effect

No discoloration

H2S, HCN

-

No effect

No discoloration

Acetylene

50 ppmr

Plus error

Discolors to white

Olefins

-

Plus error

Discolors pale yellow


Calibration gas generation Permeation tube method


TLV-TWA TLV-STEL Explosive range
2ppm - 2.0 to 12%



Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.