Chlorobenzene C6H5Cl NO.126


Performance

Measuring Range

2 to 5ppm

5 to 200ppm

200 to 500ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

0.4

1

2.5

Sampling Time

1.5 minutes per pump stroke

Detecting Limit

0.2 ppm (n=2)

Color Change

White Gray

Reaction Principle

Chlorobenzene reacts with iodine pentoxide to liberate Iodine
to discolor the reagent to gray.

Coefficient of Variation

15% (for 5 to 50 ppm), 10% (for 50 to 200 ppm)

Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Result

Change color by itself

Carbon monoxide

0.1%

No effect (2layers)

Discolors whole layer

Acetylene, Hexane, Ethylene

0.2%

No effect(2 layers)

Discolors whole layer

Alcohols, Ketones

1%

Plus error

No discoloration

Esters

0.2%

Plus error

No discoloration

Aromatic hydrocarbons

-

Plus error

Discolors to grayish brown


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
10ppm - 1.3 to 9.6%