| Chloroform | CHCl3 | NO.137 | |
|
|
|||

| Performance |
|
Measuring Range |
4 to 10 ppm |
10 to 100 ppm |
100 to 400 ppm |
|
Number of Pump Stroke |
7 |
5 |
3 |
|
Correction Factor |
0.4 |
1 |
4 |
|
Sampling Time |
2 minutes per pump stroke |
||
|
Detecting Limit |
1 ppm (n=7) |
||
|
Color Change |
White |
||
|
Reaction Principle |
Chloroform is oxidized by nascent oxygen to liberate chlorine. |
||
| Coefficient of Variation | 15% (for 10 to 20 ppm), 10% (for 20 to 100 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Humidity correction is necessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Cl2, Br2, I2 |
1/20 time |
Plus error |
Discolors to orange |
|
Unsaturated halogenated HCs |
1/20 time |
Plus error |
Discolors to orange |
|
Saturated halogenated HCs |
1/10 time |
Plus error |
Discolors to orange |
| Calibration gas generation | Diffusion tube method | |
| TLV to TWA | TLV to STEL | Explosive range |
| 10ppm | - | - |
|
|