| Carbon disulfide | CS2 | NO.13M |
|
|
||

| Performance |
|
Measuring Range |
20 to 50 ppm |
50 to 1600 ppm |
1600 to 4000 ppm |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
0.4 |
1 |
2.5 |
|
Sampling Time |
3 minutes per pump stroke |
||
|
Detecting Limit |
10 ppm (n=2) |
||
|
Color Change |
Purple |
||
|
Reaction Principle |
Carbon disulfide reacts with primary tube to produce intermediate product which changes the indicator of analyzer tube to yellow. |
||
| Coefficient of Variation | 10% (for 50 to 400 ppm), 5% (for 400 to 1600 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
| Sulfur dioxide | - |
Plus error |
Discolor yellow stain |
|
Hydrogen sulfide |
- |
Plus error |
Discolor yellow stain |
|
Lower class HCs |
1000 ppm |
Plus error |
Discolor yellow stain |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 10ppm | - | 1.3 to 50% |
|
|