Carbon disulfide CS2 NO.13M


Performance

Measuring Range

20 to 50 ppm

50 to 1600 ppm

1600 to 4000 ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

0.4

1

2.5

Sampling Time

3 minutes per pump stroke

Detecting Limit

10 ppm (n=2)

Color Change

Purple Yellow

Reaction Principle

Carbon disulfide reacts with primary tube to produce intermediate product which changes the indicator of analyzer tube to yellow.

Coefficient of Variation 10% (for 50 to 400 ppm), 5% (for 400 to 1600 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Temperature correction is necessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Sulfur dioxide -

Plus error

Discolor yellow stain

Hydrogen sulfide

-

Plus error

Discolor yellow stain

Lower class HCs

1000 ppm

Plus error

Discolor yellow stain


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
10ppm - 1.3 to 50%



Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.