Cyclohexanone C6H10O NO.154


Performance

Measuring Range

2 to 30 ppm

30 to 75 ppm

Number of Pump Strokes

4

2

Correction Factor

1

2.5

Sampling Time

2 minutes per pump stroke

Detecting Limit

0.2 ppm (n=4)

Color Change

Pale Yellow Yellow

Reaction Principle

Cyclohexanone reacts with 2,4-dinitrophenylhydrazine to form
dinitropnenylhydrazone to produce yellow in color.

Coefficient of Variation 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm)
Shelf Life 2 Years
Corrections for
temperature & humidity
Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10oC (50oF) or below.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Ketones

-

Plus error

Produce yellow stain


Other substance measurable with this detector tube

Substance

Correction Factor

Pump Strokes

Measuring Range

Furfural

1.0

4

2 to 30 ppm

Diacetone alcohol

by scale

2

2.5 to 100 ppm

Isophorone

1.0

8

2 to 30 ppm


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
25ppm - 1.1 to 9.4%



Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.