| Cyclohexanone | C6H10O | NO.154 | |
|
|
|||

| Performance |
|
Measuring Range |
2 to 30 ppm |
30 to 75 ppm |
|
Number of Pump Strokes |
4 |
2 |
|
Correction Factor |
1 |
2.5 |
|
Sampling Time |
2 minutes per pump stroke |
|
|
Detecting Limit |
0.2 ppm (n=4) |
|
|
Color Change |
Pale Yellow |
|
|
Reaction Principle |
Cyclohexanone reacts with 2,4-dinitrophenylhydrazine to form |
|
| Coefficient of Variation | 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm) | |
| Shelf Life | 2 Years | |
| Corrections for temperature & humidity |
Temperature correction is necessary | |
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | ||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Ketones |
- |
Plus error |
Produce yellow stain |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
Furfural |
1.0 |
4 |
2 to 30 ppm |
| Diacetone alcohol |
by scale |
2 |
2.5 to 100 ppm |
|
Isophorone |
1.0 |
8 |
2 to 30 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 25ppm | - | 1.1 to 9.4% |
|
|