| Carbon monoxide | CO | NO.1L |
|
|
||

| Performance |
|
Measuring Range |
2.5 to25 ppm |
25 to 1000 ppm |
1000 to 2000 ppm |
|
Number of Pump Strokes |
2 to 10 |
1 |
1/2 |
|
Correction Factor |
1/2 to 1/10 |
1 |
2 |
|
Sampling Time |
1 minute per pump stroke |
||
|
Detecting Limit |
1 ppm (n=10) |
||
|
Color Change |
Yellow |
||
|
Reaction Principle |
CO + Na2Pd(SO3)2 |
||
| Coefficient of Variation | 10% (for 25 to 300 ppm), 5% (for 300 to 1000 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Acetylene |
1/50 time |
Plus error |
Discolors dark brown |
|
Ethylene |
2000 ppm |
No effect |
Discolors whole layer to brown |
|
Hydrogen |
5000 ppm |
Plus error |
Dark brown (whole layer)(1%) |
|
Sulfur dioxide |
1/10 time |
Plus error |
Discolors blackish brown |
|
Nitrogen dioxide |
40 ppm |
Plus error |
60 ppm NO2 gives red stain |
|
Carbon disulfide |
1/50 time |
Plus error |
Discolors dark brown |
|
Halogens |
1/5 time |
Plus error |
Discolors dark brown |
|
Mercaptans |
1/5 time |
Plus error |
Discolors dark brown |
|
Hydrogen sulfide |
1/5 time |
Plus error |
Discolors dark brown |
| Calibration gas generation | High pressure gas cylinder method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 25ppm | - | 12.5 to 74% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |