| Carbon monoxide | CO | NO.1LM |
|
|
||

| Performance |
|
Measuring Range |
25 to 50 ppm |
50 to 1000 ppm |
1000 to 2000 ppm |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2 |
|
Sampling Time |
3 minutes per pump stroke |
||
|
Detecting Limit |
5 ppm (n=2) |
||
|
Color Change |
White |
||
|
Reaction Principle |
Carbon monoxide reduces iodine pentoxide to |
||
| Coefficient of Variation |
10% (for 50 to 200 ppm), 5% (for 200 to 1000 ppm) |
||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Acetylene,Olephines, |
- |
Plus error |
Produce similar stain by themselves. |
| Butane | - | Plus error | Pale Brown |
| Hydrogen | - | No | No |
| Calibration gas generation | High pressure gas cylinder method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 25ppm | - | 12.5 to 74% |
|
|