Carbonyl sulfide COS NO.21

21


Performance

Measuring Range

5 to 10 ppm

10 to 100 ppm

100 to 200 ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

1/2

1

2

Sampling Time

3 minutes per pump stroke

Detecting Limit

1 ppm (n=2)

Color Change

Blue Yellow

Reaction Formula

COS is oxidezed by iodine pentoxide and sulfuric acid to generate SO2 in primary tube. The SO2 reacts with barium chloride to libraate hydrogen chloride which discolors pH indicator (BCG) to yellow.
COS + I2O5 + H2SO4 SO2 + CO2 SO2 + BaCl2 + H2O
BaSO3+ HCl

Coefficient of Variation 10% (for 10 to 30 ppm), 5% (for 30 to 100 ppm)
Shelf Life 2 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes in the refrigerator to keep at 10oC (50oF) or below.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Carbon disulfide

-

Plus error

Produce yellow stain

Sulfur dioxide

-

Plus error

Produce yellow stain

Butane

5000 ppm

Minus error

No discoloration

Propane

5000 ppm

Minus error

No discoloration


Calibration gas generation High pressure gas cylinder method


TLV-TWA TLV-STEL Explosive range
- - 12 to 19%