Carbon dioxide CO2 NO.2LL


Performance

Measuring Range

300 to 5000 ppm

Number of Pump Strokes

300 to 5000 ppm (n=1)

Correction Factor

1

Sampling Time

2 minutes per pump stroke

Detecting Limit

30 ppm (n=1)

Color Change

White Purple

Reaction Principle

Carbon dioxide reacts with detecting agent to give indicator to purple.

Coefficient of Variation

10% (for 300 to 1000 ppm), 5% (for 1000 to 5000 ppm)

Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Ammonia

-

No effect

No discoloration

Carbon Monoxide

-

No effect

No discoloration

Hydrogen chloride,
Chlorine

-

No effect

No discoloration

Nitrogen oxides

-

No effect

No discoloration

Sulfur dioxide

-

No effect

No discoloration

Amines

-

No effect

No discoloration

Organic solvents

-

No effect

No discoloration


Calibration gas generation High pressure gas cylinder method


TLV-TWA TLV-STEL Explosive range
5000ppm 30000ppm -