| o-Cresol | C6H4(CH3)OH | NO.61 | |
|
|
|||

| Performance |
|
Measuring Range |
0.4 to 1 ppm |
1 to 25 ppm |
25 to 62.5 ppm |
|
Number of Pump Strokes |
4 |
2 |
1 |
|
Correction Factor |
0.4 |
1 |
2.5 |
|
Sampling Time |
1.5 minutes per pump stroke |
||
|
Detecting Limit |
0.1 ppm (n = 4) |
||
|
Color Change |
Pale yellow |
||
|
Reaction Principle |
Cresol react with Ceric Ammonium Nitrate to form |
||
| Coefficient of Variation |
15% (for 1 to 5 ppm), 10% (for 5 to 25 ppm) |
||
| Shelf Life | 2 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Ammonia |
2000 ppm |
Plus error |
Produce white stain |
|
Amines |
2000 ppm |
||
|
Phenol |
- |
Plus error |
Produce gray stain |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
m-Cresol |
1.0 |
2 |
1 to 25 ppm |
|
p-Cresol |
1.0 |
2 |
1 to 25 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 5ppm | - | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |