Chlorine Cl2 NO.8H


Performance

Measuring Range

25 to 50pm

50 to 500ppm

500 to 1000ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

1/2

1

2

Sampling Time

45 seconds per pump stroke

Detecting Limit

5 ppm (n=2)

Color Change

White Vermilion (reddish orange)

Reaction Principle

Chlorine is oxidized by o-Tolidine to form reddish orange color.

Coefficient of Variation 10% (for 50 to 100 ppm), 5% (for 100 to 500 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Bromine

-

Plus error

Produce vermilion

Chlorine dioxide

-

Plus error

Produce vermilion

Iodine

-

Plus error

Produce vermilion


Other substance measurable with this detector tube

Substance

Correction Factor

Pump Strokes

Measuring Range

Chlorine dioxide

0.5

1

25 to 250 ppm


Calibration gas generation Permeation tube method


TLV to TWA TLV to STEL Explosive range
0.5ppm 1ppm -


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.