Chlorine Cl2 NO.8La


Performance

Measuring Range

0.1 to 0.5ppm

0.5 to 8 ppm

8 to 16 ppm

Number of Pump Strokes

2 to 5

1

1/2

Correction Factor

1/2 to 1/5

1

2

Sampling Time

1 minutes per pump stroke

Detecting Limit

0.05 ppm (n=5)

Color Change

White Pale pink

Reaction Principle

Chlorine is oxidized by o-Toluidine to form orange products.

Coefficient of Variation 10% (for 0.5 to 2 ppm), 5% (for 2 to 8 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Nitric oxide

-

Plus error

Pale pink stain

Ozone

-

Plus error

Pale pink stain

Nitrogen dioxide

-

Plus error

Pale pink stain

Chlorine dioxide

-

Plus error

Pale pink stain

Bromine, Iodine

-

Plus error

Pale pink stain


Other substance measurable with this detector tube

Substance

Correction Factor

Pump Strokes

Measuring Range

Bromine

0.1

4

0.05 to 0.8 ppm


Calibration gas generation Permeation tube method


TLV-TWA TLV-STEL Explosive range
0.5ppm 1ppm -


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.