Chlorine Cl2 NO.8LL

8LL


Performance

Measuring Range

0.025 to 0.05ppm

0.05 to 1.0 ppm

1.0 to 2.0 ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

1/2

1

2

Sampling Time

2 minutes per pump stroke

Detecting Limit

0.005 ppm (n=2)

Color Change

White Pale Green

Reaction Principle

Chlorine oxidize the indicator to produce pale green discoloration.

Coefficient of Variation

10% (for 0.05 to 0.2 ppm), 5% (for 0.2 to 1.0 ppm)

Shelf Life 1 Year
Corrections for
temperature & humidity
Unnecessary
Store the tubes in the refrigerator to keep at 10oC (50oF) or below.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Ammonia

-

No effect

No discoloration

Carbon monoxide
Carbon dioxide

-

No effect

No discoloration

Hydrogen chloride

-

No effect

No discoloration

Sulfur dioxide

-

No effect

No discoloration

Nitrogen dioxide,
Chlorine dioxide

-

Plus error

Produce pale green

Halogens

-

Plus error

Produce pale green

Hydrogen sulfide

-

No effect

No discoloration

Organic vapors

-

No effect

No discoloration


Calibration gas generation Permeation tube method


TLV-TWA TLV-STEL Explosive range
0.5ppm 1ppm -