| 1,2-Dichloroethylene | ClCH:CHCl | NO.139 |
|
|
||

| Performance |
|
Measuring Range |
5 to 10 ppm |
10 to 100 ppm |
100 to 250ppm |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2.5 |
|
Sampling Time |
1 minute per pump stroke |
||
|
Detecting Limit |
1 ppm (n=2) |
||
|
Color Change |
Yellow |
||
|
Reaction Principle |
1,2-Dichloroethylene is decomposed by nascent oxygen by oxidizing agent to liberate hydrogen chloride which discolors indicator to reddish purple. |
||
| Coefficient of Variation | 10% (for 10 to 20 ppm), 5% (for 20 to 100 ppm) | ||
| Shelf Life | 2 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Nitrogen monoxide, Nitrogen dioxide |
- |
No effect |
No discoloration |
|
Hydrogen chloride, Halogens |
- |
Plus error |
Produce reddish purple stain |
|
Tetrachloroethylene |
- |
Plus error |
Produces reddish purple stain |
|
Trichloroethylene |
- |
Plus error |
Produces reddish purple stain |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 200ppm | - | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |