| N,N-Dimethyl acetamide | CH3CON(CH3)2 | NO.184 |
|
|
||

| Performance |
|
Measuring Range |
1.5 to 5 ppm |
5 to 60 ppm | 60 to 240 ppm |
|
Number of Pump Strokes |
4 |
2 | 1 |
|
Correction Factor |
0.3 |
1 | 4 |
|
Sampling Time |
2 minutes per pump stroke |
||
|
Detecting Limit |
0.5 ppm (n=4) |
||
|
Color Change |
Pink |
||
|
Reaction Principle |
CH3CON(CH3)2 + NaOH 2R-NH2 + H2SO4 |
||
| Coefficient of Variation | 15% (for 2 to 10 ppm), 10% (for 10 to 30 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Amines |
- |
Plus error |
Pale yellow |
|
Ammomia |
- |
Plus error |
Pale yellow |
|
Hydrazine |
- |
Plus error |
Pale yellow |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 10ppm | - | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |