| Ethanol | C2H5OH | NO.112 | |
|
|
|||

| Performance |
|
Measuring Range |
0.01 to 0.05% |
0.05 to 2.5% |
2.5 to 7.5% |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
0.2 |
1 |
3 |
|
Sampling Time |
3 minutes per pump stroke |
||
|
Detecting Limit |
0.004% (n=2) |
||
|
Color Change |
Pink |
||
|
Reaction Principle |
Ethyl alcohol reduces cromic acid to discolor pale blue stain. |
||
| Coefficient of Variation | 15% (for 0.05 to 0.5 %), 10% (for 0.5 to 2.5 %) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Carbon monoxide |
- |
No effect |
No discoloration |
|
Carbon dioxide |
- |
No effect |
No discoloration |
|
Alcohols |
- |
Plus error |
Produce similar stain |
| Calibration gas generation | Static gas dilution method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 1000ppm | - | 3.3 to 19% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |