Ethanol C2H5OH NO.112L

112L


Performance

Measuring Range

50 to 100 ppm

100 to 2000 ppm

Number of Pump Strokes

2

1

Correction Factor

1/2

1

Sampling Time

4 minutes per pump stroke

Detecting Limit

5 ppm (n=2)

Color Change

Pink Pale blue

Reaction Principle

Ethyl alcohol reduces cromic acid to discolor pale blue stain.
C2H5OH + Cr6++ H2SO4 Cr3+

Coefficient of Variation 15% (for 50 to 500 ppm), 10% (for 500 to 2000 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Temperature correction is necessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Alcohols

-

Plus error

Produce similar stain by themselves.


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
1000ppm - 3.3 to 19%


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.