| Ethylene oxide | C2H4O | NO.163LL |
|
|
|
|||

| Performance |
|
Measuring Range |
0.1 to 5 ppm |
5 to 10 ppm |
|
|
Number of Pump Stroke |
4 |
2 |
|
|
Correction Factor |
1 |
2 |
|
|
Sampling Time |
1.5 minutes per pump stroke |
||
|
Detecting Limit |
0.05 ppm (n=4) |
||
|
Color Change |
Yellow |
||
|
Reaction Formula |
C2H4O |
||
| Coefficient of Variation | 15% (for 0.1 to 0.5 ppm), 10% (for 1 to 5 ppm) | ||
| Shelf Life | 1 Year | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Aldehydes |
1/10 times |
Plus error |
Discolor pale orange |
| Ketones | - | Plus error | Discolor pale orange |
| Alcohols | 1/10 times | No effect | No discoloration |
| Calibration gas generation | Permeation tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 1ppm | - | 3.6 to 100% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |