| Formaldehyde | HCHO | NO.91L |
|
|
||

| Performance |
|
Measuring Range |
0.1 to 5 ppm |
5 to 40 ppm |
|
Number of Pump Strokes |
5 |
1 |
|
Correction Factor |
1 |
8 |
|
Sampling Time |
1.5 minutes per pump stroke |
|
|
Detecting Limit |
0.05 ppm (n=5) |
|
|
Color Change |
Yellow |
|
|
Reaction Principle |
Formaldehyde reacts with hydroxylamine phosphate to librate |
|
| Coefficient of Variation |
10% (for 0.1 to 0.5 ppm), 5% (for 0.5 to 5 ppm) |
|
| Shelf Life | 3 Years | |
| Corrections for temperature & humidity |
Temperature correction is necessary | |
| Store the tubes in the refrigerator to keep at 10oC (50oF) or below. | ||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Aldehydes |
- |
Plus error |
Produce reddish brown stain |
|
Ketones |
- |
Plus error |
Produce reddish brown stain |
|
Acid Gases |
- |
Plus error |
Produce red stain |
|
Organic acids |
- |
No effect |
No stain by themselves |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
No. of Pump strokes |
Measuring range |
|||||||
|
Diisobutyl ketone |
5.8 |
4 |
0.58 to 29 ppm |
|||||||
|
Methaldehyde |
0.65 |
3 |
0.065 to 3.25ppm |
|||||||
|
Propionaldehyde |
7.6 |
1 |
0.76 to 38 ppm |
|||||||
|
91L Tube Reading (n=1) |
0.2 |
0.5 |
1 |
2 |
3 |
4 |
5 |
|||
|
Benzaldehyde (ppm) |
2 |
4 |
9 |
22 |
40 |
63 |
92 |
|||
|
91L Tube Reading (n=1/2) |
0.2 |
0.5 |
1 |
2 |
3 |
4 |
5 |
|
Cyclohexanone (ppm) |
10 |
30 |
60 |
130 |
220 |
330 |
470 |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| - | C 0.3ppm | 7.0 to 73% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |