| Hexane | CH3(CH2)4CH3 | NO.102H | |
|
|
|||

| Performance |
|
Measuring Range |
0.015 to 0.03% |
0.03 to 0.6% |
0.6 to 1.2% |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2 |
|
Sampling Time |
1 minute per pump stroke |
||
|
Detecting Limit |
0.003% (n = 2) |
||
|
Color Change |
Orange |
||
|
Reaction Principle |
Gasoline reduces potassium dichromate to form |
||
| Coefficient of Variation | 10% (for 0.03 to 0.2 %), 5% (for 0.2 to 0.6 %) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Acetylene |
3% |
- |
Produce dark green to whole layer |
|
Propane |
0.2% |
- |
Produce dark green to whole layer |
|
Aromatic HCs |
500 ppm |
Plus error |
Produce dark green |
|
Alcohols, Esters, Ethers |
- |
Plus error |
Produce dark green |
|
Halogenated HCs. |
5000 ppm |
Plus error |
Produce dark green |
|
Hydrogen sulfide |
500 ppm |
Plus error |
Produce dark green |
|
Sulfur dioxide |
500 ppm |
Plus error |
Produce green |
| Other substance measurable with this detector tube |
|
Substance |
Correction |
No. of Pump Strokes |
Measuring Range |
|
CycloHexane |
Factor: 1.0 |
1/2, 1 or 2 |
0.015 to 1.2% |
|
Methylchyclohexane |
Factor: 1.4 |
1 |
0.04 to 0.84% |
| Calibration gas generation | Static gas dilution method | |
| TLVtoTWA | TLVtoSTEL | Explosive range |
| 50ppm | - | 1.1 to 7.5% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |