Hydrocarbons (Lower Class) C2 to C7 NO.103


Performance

Measuring Range

0.05 to 0.1%

0.1 to 1.2%

1.2 to 2.4%

Number of Pump Strokes

2

1

1/2

Correction Factor

1/2

1

2

Sampling Time

2 minutes per pump stroke

Detecting Limit

0.01% (n=2)

Color Change

Yellowish brown Greenish brown

Reaction Principle

Lower class hydrocarbons reduce chromic acid to produce greenish brown in color.

Coefficient of Variation 10% (for 0.1 to 1.2 %)
Shelf Life 2 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Carbon monoxide

-

No effect

No discoloration

Methane, Ethane

-

No effect

No discoloration

Other hydrocarbons

-

Plus error

Produce similar stain


Other substance measurable with this detector tube

Substance

Correction Factor

Pump Strokes

Measuring Range

Acetylene

1.5

1/2, 1, 2

0.075 to 3.6%

Butane, Heptane, Isobutane

0.7

1/2, 1, 2

0.035 to 1.68%

Ethylene

7.0

1/2, 1, 2

0.35 to 16.8%

Hexane

0.5

1/2, 1, 2

0.025 to 1.2%

Isopentane

0.9

1/2, 1, 2

0.045 to 2.16%

Pentane

0.75

1/2, 1, 2

0.0375 to 1.8%

Propane

1.0

1/2, 1, 2

0.05 to 2.4%


Calibration gas generation High pressure gas cylinder method


TLV-TWA TLV-STEL Explosive range
- - -