Hydrogen chloride HCl NO.14M


Performance

Measuring Range

10 to 20 ppm

20 to 500 ppm

500 to 1000 ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

1/2

1

2

Sampling Time

1 minute per pump stroke

Detecting Limit

2.5 ppm (n=2)

Color Change

Yellow Red

Reaction Principle

Hydrogen chloride reacts with indicator to produce red stain.
HCl + Indicator Chemical reaction compound

Coefficient of Variation 10% (for 20 to 100 ppm), 5% (for 100 to 500 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Hydrogen fluoride

600 ppm

Plus error

Produces red stain

Nitric Acid

1/10 time

Plus error

Produces red stain

Nitrogen dioxide

1/2 time

Plus error.

Produces red stain
(
500 ppm)


Calibration gas generation High pressure gas cylinder method


TLV-TWA TLV-STEL Explosive range
- C 5ppm -