Hydrazine N2H4 NO.185

185


Performance

Measuring Range

0.05 to 0.1 ppm

0.1 to 2.0 ppm

Number of Pump Strokes

10

5

Correction Factor

1/2

1

Sampling Time

1 minute per pump stroke

Detecting Limit

0.01 ppm (n=10)

Color Change

Pink Yellow

Reaction Principle

Hydrazine is neutralized by sulfuric acid and discolor pH indicator to yellow.

Coefficient of Variation

10% (for 0.1 to 0.5 ppm), 5% (for 0.5 to 2 ppm)

Shelf Life 3 Years
Corrections for
temperature & humidity
Humidity correction is necessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Amines & Ammonia

-

Plus error

Produces yellow stain.


Other substance measurable with this detector tube

Substance

Correction

No. of pump strokes

Measuring range

Dimethyl hydrazine

Factor:1

5

0.1 to 2 ppm

Methyl hydrazine

Factor: 6

5

0.6 to 12 ppm


Calibration gas generation Permeation tube method


TLVtoTWA TLVtoSTEL Explosive range
0.01ppm - 2.9 to 98%


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.