Hydrogen H2 NO.30


Performance

Measuring Range

0.5 to 2.0%

Number of Pump Stroke

1

Correction Factor

1

Sampling Time

2 minutes per pump stroke

Detecting Limit

0.5%(n=1)

Color Change

Yellow Yellowish brown

Reaction Principle

Hydrogen reduces sodium palladosulfide to liberate metallic palladium.

Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Special note
This is not a direct reading tube. Concentrations should be determined by using the concentration chart attached with this tube.


Possible coexisting substances and their interferences (NOTE)

Substance

Interference

Change color by itself

Carbon monoxide
Hydrogen sulfide

Produce 2 layers
(Yellowish brown and blackish brown)

Produce blackish brown

Acetylene

Produce 2 layers
(Yellowish brown and blackish brown)

Produce blackish brown

Ethylene

Plus error

Produce blackish brown


Calibration gas generation High pressure gas cylinder method


TLV-TWA TLV-STEL Explosive range
- - 4 to 75%



Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.