Hydrogen sulfide H2S NO.4LL

4LL


Performance

Measuring Range

0.25 to 2.5 ppm

2.5 to 60 ppm

60 to 120 ppm

Number of Pump Strokes

10 to 2

1

1/2

Correction Factor

1/10 to 1/2

1

2

Sampling Time

1 minute per pump stroke

Detecting Limit

0.1 ppm (n=10)

Color Change

White Brown

Reaction Principle

Hydrogen sulfide reacts with lead acetate to
form lead sulfide, which color is brown.

Coefficient of Variation 5% (for 2.5 to 60 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Nitrogen dioxide

1/4 time

Minus erorr

No discoloration

Sulfur dioxide

1 time

Plus error

No discoloration

Mercaptans,
Hydroren cyanide

-

No effect

No discoloration

Ozone - Bleaching

No discoloration


Calibration gas generation Permeation tube method


TLV-TWA TLV-STEL Explosive range
10ppm 15ppm 4.0 to 44%



Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.