Hydrogen sulfide H2S NO.4LT

4LT


Performance

Measuring Range

0.1 to 0.2 ppm

0.2 to 2.0 ppm

2.0 to 4.0 ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

1/2

1

2

Sampling Time

1 minute per pump stroke

Detecting Limit

0.05 ppm (n=2)

Color Change

Pale Yellow Reddish Brown

Reaction Principle

Hydrogen sulfide reacts with reagent to produce intermediate
substance to discolor the indicator to reddish brown.

Coefficient of Variation

10% (for 0.2 to 0.5 ppm), 5% (for 0.5 to 2 ppm)

Shelf Life 2 Years
Corrections for
temperature & humidity
Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10oC (50oF) or below.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Ammonia

-

No effect

No discoloration

Sulfur dioxide

-

No effect

No discoloration

Nitrogen dioxide

-

No effect

No discoloration

Mercaptans

1/10 time

Plus error

Produce reddish brown color

Methyl sulfide

-

No effect

No discoloration


Calibration gas generation Permeation tube method


TLV-TWA TLV-STEL Explosive range
10ppm 15ppm 4.0 to 44%


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.