Hydrogen sulfide H2S NO.4M


Performance

Measuring Range

12.5 to 25 ppm

25 to 250 ppm

250 to 500 ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

1/2

1

2

Sampling Time

30 seconds per pump stroke

Detecting Limit

2.5 ppm (n=2)

Color Change

White Brown

Reaction Principle

Hydrogen sulfide reacts with lead acetate
to form lead sulfide, which color is brown.

Coefficient of Variation 5% (for 25 to 250 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Nitrogen dioxide

1/4 timer

Minus error

No discoloration

Sulfur dioxide

1 time

Plus error

No discoloration

Mercaptans

-

No effect

No discoloration


Calibration gas generation Permeation tube method


TLV-TWA TLV-STEL Explosive range
10ppm 15ppm 4.0 to 44%


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.