| Isopropyl alcohol | CH3CH(OH)CH3 (i-C3H7OH) |
NO.113 | |
|
|
|||

| Performance |
|
Measuring Range |
0.02 to 0.04% |
0.04 to 2.5% |
2.5 to 5.0% |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2 |
|
Sampling Time |
3 minutes per pump stroke |
||
|
Detecting Limit |
0.004% (n=2) |
||
|
Color Change |
Pink |
||
|
Reaction Principle |
Isopropyl alcohol reduces potassium dichromate to form |
||
| Coefficient of Variation |
15% (for 0.04 to 0.5 %), 10% (for 0.5 to 2.5 %) |
||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Other alcohols |
- |
Plus error |
Produce similar stain by themselves |
|
Carbon dioxide |
- |
No effect |
No discoloration |
|
Carbon monoxide |
- |
No effect |
No discoloration |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
No. of pump stroke |
Measuring Range |
|
Propyl alcohol |
1 |
1 |
0.04 to 2.5 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 400ppm | 500ppm | 2 to 12.7% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |