| Isopropyl alcohol | CH3CH(OH)CH3 (i-C3H7OH) |
NO.113L |
|
|
||

| Performance |
|
Measuring Range |
25 to 50 ppm |
50 to 800 ppm |
|
Number of Pump Strokes |
2 |
1 |
|
Correction Factor |
1/2 |
1 |
|
Sampling Time |
3 minutes per pump stroke |
|
|
Detecting Limit |
5 ppm (n=2) |
|
|
Color Change |
Pink |
|
|
Reaction Principle |
Isopropyl alcohol reduces potassium dichromate to |
|
| Coefficient of Variation | 15% (for 50 to 200 ppm), 10% (for 200 to 800 ppm) | |
| Shelf Life | 3 Years | |
| Corrections for temperature & humidity |
Temperature correction is necessary | |
| Store the tubes at cool and dark place. | ||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Alcohols |
- |
Plus error |
Produce similar stain by themselves. |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
Divinyl methoxysilane |
0.05 |
2 |
2.5 to 40 ppm |
|
Ethylene glycol MEE |
1.25 |
2 |
62.5 to 1000 ppm |
|
Ethylene glycol MEEAc |
0.12 |
3 |
6 to 96 ppm |
|
1-Methoxy-2-propanol |
1.0 |
4 |
50 to 800 ppm |
|
Propyl alcohol |
1.3 |
1 |
65 to 1040 ppm |
|
Vinyl trimethoxysilane |
0.05 |
2 |
2.5 to 40 ppm |
| Ethylene glycol MBE | by scale | 2 | 30 to 1000 ppm |
| Ethylene glycol MME | by scale | 2 | 15 to 900 ppm |
|
Ethylene glycol MMEAc |
by scale | 2,4 | 20 to1300 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 400ppm | 500ppm | 2 to 12.7% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |