| Isopentyl alcohol | (CH3)2CHCH2CH2OH (i-C5H11OH) |
NO.117 |
|
|
||

| Performance |
|
Measuring Range |
5 to 300 ppm |
|
Number of Pump Strokes |
2 |
|
Correction Factor |
1 |
|
Sampling Time |
3 minutes per pump stroke |
|
Detecting Limit |
1 ppm (n=2) |
|
Color Change |
Yellow |
|
Reaction Principle |
Isoamyl alcohol reduces potassium dichromate to |
| Coefficient of Variation |
15% (for 5 to 30 ppm), 10% (for 30 to 300 ppm) |
| Shelf Life | 3 Years |
| Corrections for temperature & humidity |
Temperature correction is necessary |
| Store the tubes at cool and dark place. | |
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Alcohols |
- |
Plus error |
Produce similar stain by themselves. |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 100ppm | 125ppm | 1.2 to 9% |
|
|