1,1,1-Trichloroethane CH3CCl3 NO.135L


Performance

Measuring Range

6 to 20 ppm

20 to 200 ppm

200 to 900 ppm

Number of Pump Stroke

2

1

1/2

Correction Factor

0.3

1

4.5

Sampling Time

3 minutes per pump stroke

Detecting Limit

3 ppm (n=2)

Color Change

White Pale Pink

Reaction Formula

Methyl chloroform reacts with oxidizing agent to produce intermediate products then it reacts with detecting agent to produce pale pink stain.

Coefficient of Variation

10% (for 20 to 60 ppm), 5% (for 60 to 200 ppm)

Shelf Life 2 Years
Corrections for
temperature & humidity
Temperature correction is necessary
Store the tubes at cool and dark place.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Halogens

-

Plus error

Discolor pale pink stain

Nitrogen oxides

-

Plus error

Discolor pale pink stain

Saturated halogenated hydrocarbons

-

Plus error

Discolor pale pink stain


Other substance measurable with this detector tube

Substance

Correction Factor:

No. of pump strokes

Measuring range

1,2-Dichloroethane

5.2

1

104 to 1040 ppm

1,1,2,2-Tetrabromoethane

0.046

4

0.92 to 9.2 ppm

1,2,3-Trichloropropane

1.8

4

36 to 360 ppm


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
350ppm 450ppm -