| Methylene chloride | CH2Cl2 | NO.138 |
|
|
||

| Performance |
|
Measuring Range |
20 to 50 ppm |
50 to 500 ppm |
|
Number of Pump Stroke |
2 |
1 |
|
Correction Factor |
0.4 |
1 |
|
Sampling Time |
3 minutes per pump stroke |
|
|
Detecting Limit |
10 ppm (n=2) |
|
|
Color Change |
White |
|
|
Reaction Principle |
CH2:Cl2 + CrO3 + H2S2O7 |
|
| Coefficient of Variation |
15% (for 50 to 100 ppm), 10% (for 100 to 500 ppm) |
|
| Shelf Life | 3 Years | |
| Corrections for temperature & humidity |
Temperature correction is necessary | |
| Store the tubes at cool and dark place. | ||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Cl2, Br2, I2 |
- |
Plus error |
Discolors pale pink |
|
Unsaturated halogenated HCs |
- |
Plus error |
Discolors pale pink |
|
Saturated halogenated HCs |
- |
Plus error |
Discolors pale pink |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
Ethyl Chloride |
0.3 |
1 |
15 to 150 ppm |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 50ppm | - | 15.5 to 66.9% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |