| Methylene chloride | CH2Cl2 | NO.138L |
|
|
||

| Performance |
|
Measuring Range |
10 to 60 ppm |
60 to 150 ppm |
|
Number of Pump Stroke |
2 |
1 |
|
Correction Factor |
1 |
2.5 |
|
Sampling Time |
3 minutes per pump stroke |
|
|
Detecting Limit |
5 ppm (n=2) |
|
|
Color Change |
White |
|
|
Reaction Principle |
Methylene chloride is oxidized and produce intermediate products in the primary tube. It reacts with detecting agent to produce pale pink stain. |
|
| Coefficient of Variation | 15% (for 10 to 20 ppm), 10% (for 20 to 60 ppm) | |
| Shelf Life | 2 Years | |
| Corrections for temperature & humidity |
Temperature correction is necessary | |
| Store the tubes at cool and dark place. | ||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Cl2, Br2, I2 |
- |
Plus error |
Discolors pale pink |
|
Unsaturated halogenated HCs |
6 ppm |
Plus error |
Discolors pale pink |
|
Saturated halogenated HCs |
3 ppm |
Plus error |
Discolors pale pink |
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 50ppm | - | 15.5 to 66.9% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |