| Methyl isobutyl ketone | (CH3)2CHCH2COCH3 | NO.153 |
|
|
||

| Performance |
|
Measuring Range |
0.05 to 0.6% |
|
Number of Pump Strokes |
2 |
|
Correction Factor |
1 |
|
Sampling Time |
1.5 minutes per pump stroke |
|
Detecting Limit |
0.001% (n=2) |
|
Color Change |
Orange |
|
Reaction Principle |
(CH3)2CHCH2COCH3+ Cr6+ + H2SO4 |
| Coefficient of Variation |
10% (for 0.05 to 0.2 %), 5% (for 0.2 to 0.6 %) |
| Shelf Life | 3 Years |
| Corrections for temperature & humidity |
Temperature correction is necessary |
| Store the tubes at cool and dark place. | |
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Sulfur dioxide |
500 ppm |
Plus error |
Discolors green |
|
Hydrogen sulfide |
500 ppm |
Plus error |
Discolors dark geen |
|
Alcohols, Esters,Ketones |
- |
Plus error |
Discolors greenish brown |
|
Propane |
0.2% |
Cannot measure |
Produce dark green for whole layer |
|
Aromatic HCs |
500 ppm |
Plus error |
Discolors Greenish brown |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Stroke |
Measuring Range |
|
Styrene |
1.5 |
1 |
0.075 to 0.9% |
| Calibration gas generation | Static gas dilution method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 50ppm | 75ppm | 1.2 to 8% |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |