| Nitrogen oxides | NO+NO2 | NO.11HA | |
|
|
|||

| Performance |
|
Measuring Range |
50 to 2500 ppm |
|
Number of Pump Strokes |
1 |
|
Correction Factor |
1 |
|
Sampling Time |
1 minute per pump stroke |
|
Detecting Limit |
10 ppm (n=1) |
|
Color Change |
White |
|
Reaction Principle |
Nitrogen oxides is oxidised by nitrogen dioxide, |
| Coefficient of Variation | 10% (for 50 to 400 ppm), 5% (for 400 to 2500 ppm) |
| Shelf Life | 2 Years |
| Corrections for temperature & humidity |
Unnecessary |
| Store the tubes at cool and dark place. | |
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Hydrogen chloride |
500 ppm |
Faint demarcation |
Produce bluish purple |
|
Ozone |
200 ppm |
Faint demarcation |
Brown discoloration |
|
Sulfur dioxide, |
- |
No effect |
No discoloration |
| Calibration gas generation | NO2:Permeation tube method NO:High pressure gas cylinder method |
|
| TLV-TWA | TLV-STEL | Explosive range |
| NO2:3ppm NO:25ppm |
NO2:5ppm | - |
|
|