| Nitrogen oxides | NO+NO2 | NO.11S |
|
|
||

| Performance |
|
Measuring Range |
5 to 10 ppm |
10 to 250 ppm |
250 to 625 ppm |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2.5 |
|
Sampling Time |
1 minute per pump stroke |
||
|
Detecting Limit |
2 ppm (n=2) |
||
|
Color Change |
White |
||
|
Reaction Principle |
Nitrogen oxides is oxidised by nitrogen dioxide, |
||
| Coefficient of Variation |
10% (for 10 to 40 ppm), 5% (for 40 to 250 ppm) |
||
| Shelf Life | 2 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Hydrogen chloride |
50 ppm |
Faint demarcation |
Produce Bluish purple |
|
Ozone |
80 ppm |
Faint demarcation |
Pale brown |
|
Sulfur dioxide, |
1/1 |
Plus error |
No discoloration |
| Methanol |
400 ppm |
Minus error | No discoloration |
| Calibration gas generation | NO2:Permeation tube method NO:High pressure gas cylinder method |
|
| TLV-TWA | TLV-STEL | Explosive range |
| NO2:3ppm NO:25ppm |
NO2:5ppm | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |