| Nickel carbonyl | Ni(CO)4 | NO.20L |
|
|
|
|||

| Performance |
|
Measuring Range |
10 to 200 ppm |
200 to 800 ppm |
|
Number of Pump Stroke |
4 |
1 |
|
Correction Factor |
1 |
4 |
|
Sampling Time |
1.5 minutes per pump stroke |
|
|
Detecting Limit |
5 ppm (n=4) |
|
|
Color Change |
Pale Yellow |
|
|
Reaction Formula |
Ni(CO)4 + Gold compound |
|
| Coefficient of Variation |
10% (for 10 to 50 ppm), 5% (for 50 to 200 ppm) |
|
| Shelf Life | 3 Years | |
| Corrections for temperature & humidity |
Unnecessary | |
| Store the tubes at cool and dark place. | ||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Arsine |
- |
Plus error |
Produce purple stain |
|
Hydrogen selenide |
- |
Plus error |
Produce purple stain |
|
Hydrogen sulfide |
- |
Plus error |
Produce purple stain |
|
Phosphine |
- |
Plus error |
Produce purple stain |
|
Stibine |
- |
Plus error |
Produce purple stain |
| Calibration gas generation | Static gas dilution method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.05ppm | - | - |
|
|