| Petroleum naphtha | NO.106 | ||
|
|
|||

| Performance |
|
Measuring Range |
0.5 to 1 mg/l |
1 to 14 mg/l |
14 to 28 mg/l |
|
Number of Pump Strokes |
2 |
1 |
1/2 |
|
Correction Factor |
1/2 |
1 |
2 |
|
Sampling Time |
1 minute per pump stroke |
||
|
Detecting Limit |
0.1 mg/l (n=2) |
||
|
Color Change |
Orange |
||
|
Reaction Principle |
Petroleum naphtha reduces chromic acid to form |
||
| Coefficient of Variation | 10% (for 1 to 4 mg/l), 5% (for 4 to 14 mg/l) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Sulfur dioxide |
500 ppm |
Plus error |
Produce green color |
|
Hydrogen sulfide |
500 ppm |
Plus error |
Produce blackish brown |
|
Acetylene |
3% |
Unable to measure |
Produce blackish brown |
|
Propane |
0.2% |
Unable to measure |
Produce blackish brown |
|
Organic solvents |
- |
Plus error |
Produce greenish brown |
| Other substance measurable with this detector tube |
|
Substance |
Correction Factor |
Pump Strokes |
Measuring Range |
|
Petroleum benzene |
1.0 |
1/2, 1 or 2 |
0.5 to 28 mg/l |
|
Petroleum ether |
1.0 |
1/2, 1 or 2 |
0.5 to 28 mg/l |
| Calibration gas generation | Static gas dilution method | |
| TLV-TWA | TLV-STEL | Explosive range |
| - | - | 1.1 to 5.9% |
|
|