Pyridine C5H5N NO.182


Performance

Measuring Range

0.2 to 0.5 ppm

(0.5) to 14 ppm

14 to 35 ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

0.4

1

2.5

Sampling Time

2 minutes per pump stroke

Detecting Limit

0.1 ppm (n=2)

Color Change

Pink Yellow

Reaction Principle

C5H5N + H2SO4 C5H5N - H2SO4
Coefficient of Variation 10% (for 0.5 to 4 ppm), 5% (for 4 to 14 ppm)
Shelf Life 3 Years
Corrections for
temperature & humidity
Temperature correction is necessary
Store the tubes at cool and dark place.

Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Aromatic amines

-

Plus error

Yellow

Aliphatic amines - Plus error Yellow

Ammomia

-

Plus error

Yellow

Hydrazine

-

Plus error

Yellow

Other substance measurable with this detector tube

Substance

Correction Factor

Pump Strokes

Measuring Range

4-Methyl pyridine

0.75

1

0.38 to 10.5 ppm


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
5 ppm - 1.8 to 12.4 %


Copyright(c) 2001 GASTEC CORPORATION. All rights reserved.