| 2-Pentenenitrile | CH3CH2CH:CHCN | NO.193 |
|
|
||

| Performance |
|
Measuring Range |
0.5 to 6.0 ppm |
6.0 to 15 ppm |
|
Number of Pump Strokes |
4 |
2 |
|
Correction Factor |
1 |
2.5 |
|
Sampling Time |
3 minutes per pump stroke |
|
|
Detecting Limit |
0.1 ppm(n=4) |
|
|
Color Change |
Yellow |
|
|
Reaction Principle |
2-Pentenenitrile is decomposed by acid to liberate hydrogen cyanide which react with mercuric chloride to generate hydrogen chloride. The generated hydrogen chloride discolors indicator to red. | |
| Coefficient of Variation | 10% (for 0.5 to 2 ppm), 5% (for 2 to 6 ppm) | |
| Shelf Life | 3 Years | |
| Corrections for temperature & humidity |
Temperature correction is necessary |
|
| Store the tubes at cool and dark place. |
||
| Possible coexisting substances and their interferences |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Hydrogen chloride |
- |
No effect |
No discoloration |
|
Hydrogen cyanide |
- |
No effect |
No discoloration |
|
Nitriles ( |
- |
Plus error |
Discolor red stain |
|
Acetone cyanohidrin |
- |
Plus error |
Discolors red stain |
|
Alcohols, Esters, Ketones |
20 ppm or higher |
Minus error |
No discoloration |
|
Aromatic hydrocarbons |
20 ppm or higher |
Minus error |
No discoloration |
Note: Humidity scrubber removes humidity and acid gases.
| Calibration gas generation | Diffusion tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| - | - | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |