| Phosphine | PH3 | NO.7J | |
|
|
|||

| Performance |
|
Measuring Range |
2.5 to 25 ppm |
25 to 500 ppm |
500 to 1000 ppm |
|
Pump Strokes |
2 to 10 |
1 |
1/2 |
|
Correction Factor |
1/2 to 1/10 |
1 |
2 |
|
Sampling Time |
1 minute per pump stroke |
||
|
Detecting Limit |
1 ppm (n = 10) |
||
|
Color Change |
White |
||
|
Reaction Principle |
Phosphine reacts with mercuric chloride to produce |
||
| Coefficient of Variation |
10% (for 2.5 to 100 ppm), 5% (for 100 to 500 ppm) |
||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Arsine |
1/3 time |
Plus error |
Produces brown stain |
|
Hydrogen selenide |
1/5 time |
Plus error |
Produces brown stain |
|
Hydrogen sulfide |
1/5 time |
Plus error |
Produces dark brown stain |
| Calibration gas generation | High pressure gas cylinder method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.3ppm | 1ppm | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |