| Sulfur dioxide | SO2 | NO.5L | |
|
|
|||

| Performance |
|
Measuring Range |
1.25 to 2.5ppm |
2.5 to 5ppm |
5 to 100ppm |
100 to 200ppm |
|
Number of Pump Strokes |
4 |
2 |
1 |
1/2 |
|
Correction Factor |
1/4 |
1/2 |
1 |
2 |
|
Sampling Time |
1 minute per pump stroke |
|||
|
Detecting Limit |
0.25ppm (n=4) |
|||
|
Color Change |
Blue |
|||
|
Reaction Principle |
Sulfur dioxide reacts with barium chloride to generate |
|||
| Coefficient of Variation |
10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm) |
|||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity |
Unnecessary | |||
| Store the tubes at cool and dark place. | ||||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Carbon monoxide, |
- |
No effect |
No discoloration |
|
Carbon dioxide |
- |
No effect |
No discoloration |
|
Nitrogen dioxide |
1:1 |
Plus error |
Produces light purple discoloration |
|
Hydrogen sulfide |
- |
No effect |
No discoloration |
| Calibration gas generation | Permeation tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 2ppm | 5ppm | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |