| Sulfur dioxide | SO2 | NO.5LC | |
|
|
|||

| Performance |
|
Measuring Range |
0.1 to 0.25 ppm | (0.25) to 10 ppm | 10 to 25 ppm |
|
Number of Pump Strokes |
4 | 2 | 1 |
|
Correction Factor |
0.4 | 1 | 2.5 |
|
Sampling Time |
2 minutes per pump stroke |
||
|
Detecting Limit |
0.02 ppm (n=4) |
||
|
Color Change |
Bluish purple |
||
|
Reaction Principle |
Sulfur dioxide reacts with reaction agent to produce |
||
| Coefficient of Variation | 10% (for 0.25 to 2 ppm), 5% (for 2 to 10 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity |
Temperature correction is necessary | ||
| Store the tubes at cool and dark place. | |||
| Possible coexisting substances and their interferences | (NOTE) |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Nitrogen dioxide |
- |
Minus error |
Produces brown stain of 1 mm |
|
Hydrogen sulfide |
1/10 |
Plus error |
Produce unclear demarcation |
| Calibration gas generation | Permeation tube method | |
| TLV-TWA | TLV-STEL | Explosive range |
| 2ppm | 5ppm | - |
|
|
| Copyright(c) 2001 GASTEC CORPORATION. All rights reserved. |